As an example, the data for a maize analysis project might include all the measured spectra with information on the variety, harvest year, region of origin or other significant variables and a LIMS identifier linking each sample with analysis data such as dry-matter or protein. All the developed calibration models together with the resulting tested methods for dry-matter and protein prediction are documented together with validated and released methods for single or multi-component analysis. Harmonised Workflow The CPF file provides the common platform for the individual SL programs, so that for example SL Predictor can be run on the lab computer which is connected to the spectrometer while SL Calibration Workshop is in use on a PC in the office. The first step in calibration development is to capture the sample spectra with SL Predictor or to use previously recorded data from a source such as JCAMP or SPC. The sample data are imported into the project file. Next, SL Calibration Workshop is used to generate the calibration models. This procedure uses well-proven tools and procedures and is focussed on the essential operations so as to keep the user time requirement to a minimum. For quantitative analysis a multi-linear regression can be applied directly to the wavelength variables (MLR) of the spectra or it can be applied to the derived factors from a variance analysis using Principal Component Regression (PCR) or Partial Least Squares Regression (PLSR). For qualitative analysis a discriminant module for spectral libraries and a cluster model are available, both with factoring by Principal Component Analysis (PCA). Calibrations which appear to be suitable are then incorporated into a Method which is tested by measuring an independent set of samples not included in the calibration and, if successful, may subsequently be validated. Three freely selectable combinations from validated methods can be combined and released for routine analysis so that for example a quantitative application for dry-matter measurement entitled “Maize” which has been proven during one year’s harvest may be extended by adding calibration samples from the new crop and also by adding the sample data for a further measurement such as protein. The project file containing the Maize application can then be used for direct measurement using SL Predictor. Routine Analysis Before the first measurement a few simple steps of preparation are made so as to simplify the daily work; for example, the user selects the established application, or calibration, to be used for routine analysis and, if wanted, the option to add the analysis values for other parameters if they will already be available for the samples. Figure 2 shows an example, with an additional “Quality Index”.
After the sample has been measured the application automatically checks whether the sample is valid for the calibration being used and marks the result accordingly in green or red (Figure 3). In the case of red, a code indicates which aspect of the outlier check failed:
As well as checking outliers the software provides an estimate of the measurement error. This indicates the limits within which a reference analysis would have a 95% probability of agreeing with the measured result. SL Predictor calculates this according to the method of ASTM E1655, part 15.4.1. Results from a qualitative analysis are presented with similar clarity. For a conformity analysis (“Is this sample the same as X or not?”) a positively identified sample is marked as OK. If the result is uncertain, the program shows the user a list of substances with the most similar spectra from the selected library. Results can be printed in report form or sent as an ASCII file to a LIMS. They are also stored in the CPF project file. Spectrometer Connections Wherever possible the SL software incorporates the original drivers from a wide range of spectrometer manufacturers so as to ensure a seamless connection to the spectrometer and without tying the user to a particular manufacturer or type. All the necessary operating functions can be carried out from the SL Predictor user interface (Figure 4).
In close co-operation with spectrometer manufacturers and in accordance with their recommendations, all instrument-specific settings and procedures are implemented, for example for a single-beam spectrometer a reference measurement can be performed with simultaneous optimisation of the signal level. SL Predictor adds other useful functions for all instrument types, such as a baseline measurement to correct the extinction value for a particular sample container, or other type of interference which would not be corrected in a reference measurement. This provides the user with a program which is ready to use, so that during installation it is only necessary to select the instrument driver(s) to be loaded: the system is then ready for use with the selected spectrometer. Accessories In addition to the main applications the SL software family offers a range of useful accessory programs. These are bundled together in the SL Utilities. There are tools to support the direct or indirect import of existing data in a CPF database, for example the batch import of individual SPC spectra, optionally combined with the associated reference analysis values in an Excel file. The “Subset Selection“ allows representative spectra sets for calibration and validation data sets to be assembled by selecting significant spectra, either with the help of a Gauss-Jordan procedure [3] or at random. SL Database Viewer (Figure 1) provides an independent tool for quickly viewing and managing CPF project files, and the SL Application Development Kit offers the possibility to read from and write to CPF files from other software programs and to use the analysis methods they contain. Summary The user in our original example for Maize analysis can – provided spectrometer manufacturers A and B have provided their drivers - now perform his dry-matter analysis as easily as this:
In this way all the procedures are brought together in a simple and easy to use way, so that the work becomes more efficient and the risk of errors is reduced, while at the same time meeting the requirements of quality assurance for the chemometric analysis. References [1] T. Nĉs, T. Isaksson, T. Fearn, T. Davies: [2] H. Mark: [3] D.E. Honigs, G.M. Hieftje, H.L. Mark, T.B. Hirschfeld: With grateful thanks to KWS Saat AG / Einbeck for permission to publish maize spectra in the application example. Additional information
Products
|
|
|||||||||||||||||||||||||||||||||||||||||||||||||||
| Product of the Week |
||||||||||||||||||||||||||||||||||||||||||||||||||||